Harmonization of growth hormone measurements with different immunoassays by data adjustment

Abstract:

BACKGROUND\backslashr\backslashnThe aim of our study was to evaluate the between-assay variability of commercially available immunoassays for the measurement of human growth hormone (hGH). In addition, we asked whether the comparability of the diagnosis of childhood onset growth hormone deficiency could be improved by adjusting hGH results by statistical methods, such as linear regression, conversion factors, and quantile transformation.\backslashr\backslashnMETHODS\backslashr\backslashnIn archived sera from 312 children and adolescents (age: 17 days-17 years) hGH values between 0.01 and 16.5 ng/mL were determined by using the following immunoassays: AutoDELFIA (PerkinElmer), BC-IRMA (Beckman-Coulter), ELISA (Mediagnost), IMMULITE 2000 (Siemens), iSYS (IDS), Liaison (DiaSorin), UniCel DxI 800 Access (BeckmanCoulter) and {\textquotedblIn house{\textquotedbl-RIA (Tübingen).\backslashr\backslashnRESULTS\backslashr\backslashnThe assays differed in median hGH concentrations by as much as 5.44 ng/mL (Immulite), and as little as 2.67 ng/mL (BC-IRMA). The mean difference between assays ranged from 0.35 to 2.71 ng/mL, whereas several samples displayed differences up to 11.4 ng/mL. The best correlation (r=0.992) was found between AutoDELFIA and Liasion, the lowest (r=0.864) was between an in-house RIA and iSYS. The between-assay CV (mean \pm SD) of values within the cut-off range was 24.3% \pm 7.4%, resulting in an assay-dependent diagnosis of growth hormone deficiency (GHD) in more than 27% of patients. Yet, adjustment of this data by linear regression or a conversion factor reduced the CV below 14%, and the ratio of assay-dependent diagnoses below 8%. Using quantile transformation, the CV and ratio were reduced to 11.4% and \textless1%, respectively.\backslashr\backslashnCONCLUSIONS\backslashr\backslashnhGH measurements using different assays vary significantly. Linear regression, conversion factors, or particularly quantile transformation are useful tools to improve comparability in the diagnostic procedure for the confirmation of GHD in childhood and adolescence.

DOI: 10.1515/CCLM.2011.201

Projects: Genetical Statistics and Systems Biology

Publication type: Journal article

Journal: Clinical chemistry and laboratory medicine : CCLM / FESCC

Human Diseases: No Human Disease specified

Citation: Clinical Chemistry and Laboratory Medicine 49(7)

Date Published: 2011

Registered Mode: imported from a bibtex file

Authors: Anne Müller, Markus Scholz, Oliver Blankenstein, Gerhard Binder, Roland Pfäffle, Antje Körner, Wieland Kiess, Annegret Heider, Martin Bidlingmaier, Joachim Thiery, Jürgen Kratzsch

Help
help Submitter
Citation
Müller, A., Scholz, M., Blankenstein, O., Binder, G., Pfäffle, R., Körner, A., Kiess, W., Heider, A., Bidlingmaier, M., Thiery, J., & Kratzsch, J. (2011). Harmonization of growth hormone measurements with different immunoassays by data adjustment. In Clinical Chemistry and Laboratory Medicine (Vol. 49, Issue 7). Walter de Gruyter GmbH. https://doi.org/10.1515/cclm.2011.201
Activity

Views: 1787

Created: 14th Sep 2020 at 13:13

Last updated: 7th Dec 2021 at 17:58

help Tags

This item has not yet been tagged.

help Attributions

None

Related items

Powered by
(v.1.13.0-master)
Copyright © 2008 - 2021 The University of Manchester and HITS gGmbH
Institute for Medical Informatics, Statistics and Epidemiology, University of Leipzig

By continuing to use this site you agree to the use of cookies